Digital I/O Channel Numbering Rules:

  1. The Digital I/O channel numbering system is based on module codes. In a synchronized multichannel AE diagnostic system formed from AED404 and/or AED416 devices, all Digital I/O channels have different channel number.
  2. All AED416 devices have 8 Digital I/O channels. For any of these devices the starting Digital I/O channel number is: (Module_CodeAED416 – 1)*16 + 1.
  3. All AED404 devices have 3 Digital I/O channels. For any of these devices the starting Digital I/O channel number is: (Module_CodeAED404 – 1)*4 + 1.