Module coding rules:
-
The EMA-4/16 devices can be set with any module code from 1 to 8. All EMA-4/16 devices in a synchronised AE diagnostic system must have different module code.
-
The EMA-4/4 devices can be set with any module code from 1 to 32. All EMA-4/4 devices in a synchronised AE diagnostic system must have different module code.
-
The AE channel numbering system is based on module codes. In a synchronised multichannel AE diagnostic system formed from EMA-4/4 and/or EMA-4/16 devices,
all AE channels must have different channel number.
-
All EMA-4/16 devices have 16 acoustic-emission channels. For any of these devices the starting AE channel number is: (Module_CodeEMA-4/16 – 1)*16 + 1.
-
All EMA-4/4 devices have 4 acoustic-emission channels. For any of these devices the starting AE channel number is: (Module_CodeEMA-4/4 – 1)*4 + 1.